The FAQs below are strictly referred to the products from Norditalia and are not valid for products of other manufacturers. It is mandatory to read carefully the user manual that comes with the appliance. The informations contained in this section are complementary and not substitutive to the contents of the user manual.
Traditional Aerosol Therapy Devices.
Before using these appliances is of utmost importance to read carefully the user manual for the product.
If the case the device is not working, please check this out:

Ultrasound Device MO-03.
Before using the device is of utmost importance to read carefully the user manual of the product.
In the case the aerosol therapy appliance is not working, please check this out:
WARNING: the appliance must always be used by filling the water cup (7) with normal water to the level specified in the user manual. Put the medicine into the medine cup (5) and place it inside the device. Never put the medicine directly into the water cup (7) without using the medicine cup. This way the device is irreparably damaged.

ACCESSORIES CONNECTION SCHEME

PROBLEM/CAUSE/SOLUTION SUMMARY TABLE

| Problem | Display Indication |
Cause | Solutions |
| The appliance switches itself off or does not come on | |
No water or water below the reference level | Fill the reservoir suitably |
| When using the appliance the battery symbol flashes | The battery is being charged | Switch off the appliance and charge the battery | |
| If you are using the adapter it indicates a temporary lowering of the power supply voltage | You can continue with the therapy, unless the nebulization becomes insufficient | ||
| The appliance does not come on and the battery symbol flashes | |
Flat battery | Charge the battery or power the appliance with the adapter |
| The appliance switches itself off after around 18 minutes. The timer symbol comes on and stays on | |
Timer intervention | Let the appliance cool for at least 30 minutes |
| The appliance switches itself off and the timer symbol flashes | |
The appliance is overheated | Let the nebulizer cool for at least 30 minutes |
| The appliance remains switched off after pressing ON/OFF | Green LED switched off | Adapter not connected correctly | Check that the adapter is well connected both to the nebulizer and to the electrical socket |
| Battery not connected correctly | Make correct connection. See the user manual. | ||
| Appliance/battery/adapter damaged | Contact the stockist or an assistance center | ||
| Nebulisation very poor or absent | Too much medicine (more than 8 ml.) in the medicine cup | As indicated in the "USE INSTRUCTION" subdivide the dose | |
| Small amount of medicine | As indicated in the "use instructions" do not insert less than 4 ml. | ||
| Medicine not inserted | Check that medicine is inserted in the medicine cup | ||
| Medicine not suitable for use with ultrasound appliances | Check the instructions enclosed to the medicine and consult a doctor | ||
| Small amount of water in the reservoir | Check that the water arrives at the level indicated | ||
| Fan blocked by encrustation | See cleaning instructions | ||
| Wet filter / dirty filter | Rinse the filter with water and dry it with a cloth / if necessary replace the filter | ||
| Noisy appliance | Noisy fan | Check that the fan is not encrusted with medicine. Check that the filter does not touch the fan | |
| Sudden nebulisation interruption | Air bubble between the cover and the tube | Blow gently into the tube | |
| Wet filter | Rinse the filter with water and dry it with a cloth | ||
| Medicine has escaped from the medicine cup due to brusque movements |
Replace the medicine |